Contents |
Generally any Cryostat should be appropriate for 10 – 50μm sectioning of frozen soft tissue onto glass slides at temperatures between -10 and -40 degrees Celsius. These guideline parameters can be exceeded with discretion. Read the users manual before using the cryostat!
Before you begin sectioning, you should check the following:
All materials (besides the glass slides and OCT compound) should be in put in the cryostat for at least 2 hours before starting to cut. In particular, any tissue to be cut, chucks to be mounted upon, blades to be cut with, and brushes for the sweepin.
You can usually turn on the instrument by pressing any of the buttons or opening the chamber window (the latter is preferred), the display shows might show something like:
BUTTON FUNCTIONS For additional info please refer to the users manual.
UP/DOWN buttons These two buttons are used to change pre-selected values of different kinds in connection with the chosen function.
Object temperature button This button allows you to set the object quick freeze for the unused chucks
Chamber temperature button This button allows you to set the chamber temperature using the UP/DOWN buttons.
Fine and Trim thickness button This button highlights and toggles between the fine thickness (the thickness you actually cut sections at) and trim thickness (the thickness a section is cut after pressing the trim button 1 time). These values are changed by small preset intervals using the UP/DOWN buttons.
Reset button This button is used to reset added values (e.g. number of sections, sum of section thicknesses) to zero.
Menu button This button is used to get into the submenus. Basic settings, times, service functions and options are set and/or called.
Enter/scroll button Within the submenu, basic settings are carried out, further submenus are called and other functions are activated via this button. During operation of the instrument, e.g. while cutting, this button is used to choose between the indication of the section thickness, section sum, real time and indication of the distance to the end position of the specimen.
SPECIMEN COARSE FEED For the fast forward and backward travel between knife and specimen, the cryostat microtome has a motorized coarse feed system. The approach between specimen and knife can also be carried out by means of an automatic function
Chuck Return travel button
Chuck Feed Travel Button
Cutting Window Button To activate an automatic approach process, it is necessary to set a cutting window first. The size of this window is then shown in the symbol cutting window. When passing the cutting window this symbol is shown inverted on the display. Manual setting:
Note: A cutting window should only be set while the specimen is moved downwards. If, by mistake, a cutting window limit is set during return travel of the specimen, the set limits are applied to the cutting movement accordingly. Note: In case the cutting window limits are set incorrectly, please repeat above-mentioned process again. Note: If no cutting window is needed, briefly press the Cutting Window Button twice. The automatic approach system can now be released in any position.
Automatic Chuck Approach Button
SELECTION CUTTING MODES For the motorized cutting movement of the microtome, the following operating modes are available: - continuous stroke - single stroke - interval stroke
Cutting Mode Button
Continuous stroke (1)
Single stroke (2)
Interval stroke (3)
SETTING THE CUTTING SPEED
Cutting Speed Knob
Note: To save time, the return travel speed is enhanced in relation to the cutting speed, especially for slow cutting speeds.
START AND STOP OF THE CUTTING DRIVE
Cutting Start/Stop Buttons
Note: To start the motorized cutting drive, the functions handwheel brake and emergency stop must not be activated. The course function of the cutting drive results from the selected cutting window, the selected operating mode and the set cutting speed.
Stop in operating mode continuous stroke
Handwheel Brake Button Unintended movements of the specimen holder can be avoided via the handwheel brake. This reduces the danger of being injured while adjusting specimen clamp and knife carrier! Warning: When the instrument is turned off, the handwheel brake is not operative. For your personal safety, the handwheel brake should be turned on when working on the specimen holder or knife carrier.
SETTING THE STAGE VACUUM FOR DISPOSING OF SECTION WASTE
Setting the vacuum
Note: The user sets the desired value depending on the desired function (stretching or suction process), section thickness and size of the specimen. Note: The applied vacuum for stretching sections and/or disposing of section waste is only active during the cutting window representing also the suction window. Outside the cutting window, the selected vacuum is turned off by means of a valve. Setting the suction window is carried out in the same way as the cutting window.
WHEN THE VACUUM FILTER IS FULL, THE CRYO WILL TELL YOU THAT IT NEEDS TO BE CHANGED.
INSERTING AND REPLACING THE FILTER UNIT When the suction power and suction window are set in a correct way, the section to be disposed of gets via the suction slot in the knife carrier into the filter unit, consisting of a coarse filter (fig. 93-2) for collecting section waste and a micro filter for filtering the air. Section waste is collected in the coarse filter. This coarse filter must be replaced depending on the volume of waste. The coarse filter is actively cooled. The micro filter must also be replaced regularly, however, at least every 6 months. Removing coarse filter:
Replacing the coarse filter: The coarse filter must be replaced depending on the volume of the existing section waste. If the integrated counter detects that the sum of all sections exceeds 50,000 μm after the last resetting, the message "Change coarse filt." appears on the display.
Note: The coarse filter is to be understood as a disposable filter and must not be used again.
WHEN YOU ARE DONE CUTTING IT IS TIME TO CLEAN UP!
Clean the cryo chamber from all section waste after.
Thoroughly Clean the Chamber!! Remove frozen section waste with a cold brush. Empty the section waste tray. Wipe down the storage shelves and brush shelf with 75% ethanol. The cryostat should be left immaculately clean after each use. Remove all specimens from the cryostat. Close the sliding window.
EMPTYING THE DEFROSTING LIQUID
Note:
The liquid from the daily defrosting process is collected in the container (fig. 62.1). Please note that this container is emptied in due time according to the safety precautions at the beginning of this instruction manual. We highly recommend to treat the container with a disinfection solution after having emptied it.
Dec 18 2011 4:24 am
Good tip. I did it and it worked. Thanks.
Jan 05 2012 12:32 am
Very nice post. I just stumbled upon your weblog and wished to say that I've truly enjoyed browsing your blog posts. In any case I will be subscribing to your rss feed and I hope you write again soon!
Jan 12 2012 9:27 pm
Sorry, but I disagree with this unique story. I do take pleasure in your web page still and may keep on moving again for upgrades.
Apr 17 2012 4:08 pm
Nice post. I learn one thing more challenging on totally different blogs everyday. It'll always be stimulating to read content material from other writers and observe slightly something from their store. I’d desire to make use of some with the content on my weblog whether you don’t mind. Natually I’ll provide you with a link on your web blog. Thanks for sharing.